Welcome back. The previous lesson separated a reading’s location from its relevance to another place. Two instruments can also disagree while sampling the same air. Here you will examine one reason: a low-cost particulate-matter sensor usually detects light scattered by particles, then estimates mass. Humidity and the kinds of particles present can change that estimate.
By the end, you should be able to explain why a humid morning or a change from smoke to dust might alter a sensor’s reported PM2.5 even when the change in dry particle mass is smaller—or goes in a different direction.
How light becomes a PM reading
PM2.5 is reported as a mass concentration, usually in . A typical low-cost optical sensor does not weigh the particles in a cubic meter of air. It draws in a sample, shines light through it, and detects light scattered by particles. Depending on the design, it measures scattering from many particles together or estimates the sizes and numbers of individual particles. Software converts that optical signal into a reported mass concentration.
That conversion requires assumptions. A particle’s scattering depends not just on how much material it contains, but also on its size, shape, and optical properties. Estimating mass additionally requires assumptions about density and the distribution of particle sizes. In a simplified model of a spherical particle, volume is proportional to the cube of its diameter, . A modest error in estimated size can therefore make a substantial difference to estimated mass. Real sensor algorithms are more complicated, but the essential limitation remains: light scattering is evidence about particles, not a direct measurement of their dry mass.
Read the World Meteorological Organization’s Low-cost sensors for the measurement of atmospheric composition for the measurement principle and its comparison with more controlled instruments.
Low-cost sensors for the measurement of atmospheric composition
The World Meteorological Organization’s review explains what optical PM sensors measure and why moisture and particle type affect the conversion to mass.
In Section 3.2, “Low-cost sensors for Particulate Matter (PM)” (printed pp. 19–20), begin with the paragraph starting “Particle measurements, when categorized across different sizes.” Read the operating principle. Later in the same section, find the paragraph beginning “The most significant interfering variable seen with low-cost PM sensors” and read the humidity discussion. Finally, on printed p. 20, find the comparison paragraph beginning “Though they are often based on similar analytical principles” and read the inlet comparison. Focus on the difference between sensing particles at ambient humidity and conditioning the sample before measurement.
The review also notes that small particles can fall below a low-cost sensor’s detection limit. A sensor’s estimate can thus depend on which particle sizes are present, even before humidity enters the picture. This does not make every low-cost reading useless; it means the reported number must be understood as a calibrated estimate made by a particular instrument.
Why moisture changes what the sensor sees
Relative humidity describes how close the air is to saturation with water vapor at its current temperature. Water vapor itself is not a PM2.5 particle. The important interaction is with airborne particles.
Some particles are hygroscopic: they take up water as humidity rises. Salts are a familiar example. Once a suitable humidity is reached, a salt-containing particle can become a larger, wetter particle. It may scatter light differently, and an optical sensor may assign it a larger size or a higher PM mass estimate. Near saturation, suspended mist or fog droplets can also scatter light and be detected as particles. These are two related but distinct effects: water added to existing particles and additional liquid droplets in the air.
Many low-cost sensors sample particles at ambient humidity. Some reference measurement systems control humidity or dry the sample to make measurements closer to a defined, dry-equivalent PM concentration. Consequently, two instruments can respond differently to the same humid air without either display necessarily being fabricated. A high uncorrected optical reading can reflect real wet particles or droplets in the air while still being a poor estimate of the dry-equivalent concentration used for a particular air-quality comparison.
The research article The influence of humidity on the performance of a low-cost air particle mass sensor and the effect of atmospheric fog connects this mechanism to measurements. Its laboratory result is especially useful because the particle concentration supplied to the chamber was kept approximately steady while humidity changed.
This Atmospheric Measurement Techniques article shows how wet particle growth and fog affected a particular optical sensor, providing a concrete test of the mechanism.
In the introductory discussion, start at “Humid conditions can affect the performance of a sensor” and read the mechanism. Then find the laboratory-chamber paragraph immediately before Section 3.2, “Field experiments”; it begins “With the steady introduction of ambient air.” Read the chamber result, using Figure 2 to follow the humidity change. In Section 3.2, find the paragraph beginning “Figure 4 shows the hourly PM2.5 concentrations” and read the fog example. Compare the instruments’ trends rather than treating either single reading as a universal correction factor.
In that laboratory test, the sensor’s indicated PM2.5 rose from about to as relative humidity increased from about to , although the supplied particle concentration was held approximately steady. The study linked much of the rise to particles taking up water. In a foggy field period, the optical sensor rose sharply while the comparison instrument changed much less.
These results do not establish a universal humidity threshold. The article found an effect beginning around – for the conditions it studied; the WMO review describes degradation often seen above roughly –. The response depends on the particles, the instrument, and recent conditions. High humidity also does not prove that a pollution increase is false: dry particle concentrations can genuinely change at the same time.
Why particle composition matters even without fog
Consider two air samples with the same dry PM2.5 mass. One contains relatively dense mineral dust; the other contains smaller smoke particles. Their particle counts, sizes, densities, shapes, and optical properties can differ. The same mass need not scatter the same amount of light. If a sensor was calibrated using one mixture, its estimate for the other may be too high or too low.
Composition also determines the humidity response. Salt-containing particles can absorb substantial water under suitable conditions; another mixture may grow much less at the same relative humidity. Thus, “the humidity was ” is not enough to calculate how far a PM reading is biased. Nor is one correction necessarily transferable from urban aerosol to wildfire smoke or from one season to another.
The diagram below sketches why a research calibration might account for both optical response and aerosol liquid water—often abbreviated ALW. Treat it as an illustration of factors that can be considered, not as a correction performed automatically by every air-quality service.

A useful distinction is between an optical estimate of particles as sampled and a dry-equivalent PM estimate. Removing or estimating the water contribution is not as simple as subtracting a fixed number whenever humidity is high. It may require a tested correction and information about the aerosol mixture. Conversely, drying and reference methods have their own measurement limitations; “reference” does not mean unaffected by every condition.
Interpreting a disagreement
Suppose a nearby low-cost sensor reports rising PM2.5 before dawn while a comparison monitor stays fairly steady. If humidity is climbing and fog is present, wet particle growth or droplets are plausible contributors to the difference. If the readings separate during a shift from smoke to dust, a change in particle-size distribution or optical properties is another plausible contributor. Neither explanation can be established from the two PM numbers alone.
For this instrument-response question, the most informative context is the humidity and fog conditions, the likely particle source, whether the reported value is raw or corrected, and whether the instruments condition their samples differently. This complements the previous lesson’s location check: instruments may disagree because they sampled different air, because they responded differently to the same air, or both. The next lesson will develop the broader reliability assessment, including calibration evidence, data completeness, and quality-control flags.
Key takeaways
Low-cost optical PM sensors infer mass from scattered light rather than directly weighing dry particles. Humidity can make hygroscopic particles grow and can introduce detectable fog droplets, often raising an uncorrected estimate relative to a dry-equivalent measurement. Particle composition changes both how strongly particles scatter light and how much water they absorb, so the size—and sometimes the direction—of a sensor’s error cannot be read from humidity alone. These mechanisms explain possible disagreements; they do not, by themselves, prove which reported concentration is most reliable.